DM Series GC Capillary Column Installation Instructions
1. Check Instrument System
- Cool all heated zones, such as injection, detector and so on.
- Visually inspect septum, liner, O-ring, and replace if broken or dirty.
2. Prepare Column for Installation
- Set make-up and detector gas flow rates.
- Install a nut and an appropriately sized ferrule on each end of column.
- Cut 1–2 cm from each end of column to prevent ferrule fragments. Port cutting must be smooth.
- Place the column on the holder in the oven, keep the side with words towards your face.
3. Install Column
- Connect the column to the inlet at the instrument manufacturer’s recommended insertion distance.
- Set column head pressure or column flow rate, split ratio, septum purge and any other applicable inlet parameters.
- Confirm flow by immersing column outlet in a vial of organic solvent (e.g., n-hexane). Observe whether there are uniform air bubbles. If yes, clean the solvent from the column outlet and cut 1–2 cm from the column end. If no, check the instrument system.
- Connect the column to the outlet at the instrument manufacturer’s recommended insertion distance.
Attention:
Some detectors (e.g., ECD, NPD, MS) specify that a new
column be conditioned before being connected to the
detector. Also, detector must be sealed with no-hole
ferrule and nut.
4. Check for Leaks
- Check for leaks at the injector and detector, using a thermal conductivity leak detector. Do not use soaps or liquids to check for leaks.
- Set injector and detector temperatures and turn on detector.
Attention:
Do not exceed the phase’s maximum isotherm operating
temperature! See the attached temperature limits list.
- Check again for leaks, because O2 will damage the column.
5. Conditioning Column
To further ensure the homogeneity and stability of phase, the new column must be conditioned before using. This helps provide a more stable baseline.
- Before conditioning, purge the column at 40–50 ℃ for a minimum of 20 min.
Normal Capillary Column Conditioning Method
Normal Capillary Column
Initial temperature:
50 ℃
Heating rate: 5 ℃/min
Raise the temperature to the conditioning temperature, which is about 20 ℃ higher than the maximum experiment using temperature.
Never exceed the phase’s maximum isotherm operating temperature. The suggested temperature is 10–20 ℃ lower than the phase’s maximum isotherm operating temperature.
Keep at the conditioning temperature for 1–2 h.
PLOT Series Capillary Columns
PLOT series capillary column conditioning methods are different because of the phase.
a. PLOT Q, QS, S, U Column
Initial temperature:
50 ℃
Heating rate: 5 ℃/min
Raise the temperature to the phase’s maximum isotherm operating temperature.
Keep for at least 1 h.
b. PLOT MS 5A, Alumina, Alumina/Na2SO4,
Alumina/KCl, CFC Column
Initial temperature:
50 ℃
Heating rate: 5 ℃/min
Raise the temperature to the phase’s maximum isotherm operating temperature.
Keep for at least 2 h.
- For sensitive detectors or highly sensitive analyses, longer conditioning times might be necessary until the baseline is stable. Conditioning time should not be more than 4 h for any capillary column.
- When using Vespel or graphite/Vespel ferrules, re-tighten the fitting after conditioning, and periodically check for leaks.
- If the column is not connected to the detector (e.g., ECD, NPD, MS), cut 10–20 cm from the end of column, make the connection with detector, and check for leaks.
- If necessary, follow the COA method to evaluate whether the column is qualified.
- Please check the instrument system again when emerging questions, such as tailing peaks or poor reproducibility.
- Set instrument parameters and start the experiment.
Temperature Limits:
Each column has temperature lower and upper limits.
If exceeding the temperature lower limit, resolution
and peak shape will be abnormal, but this will not
damage the column.
The temperature upper limit consists of two temperatures. The lower temperature is the maximum isotherm temperature on which the column can be used normally. The higher temperature is the maximum programming temperature and should not be held for more than 10 min.
Operating at a higher temperature than the temperature limit will reduce the column lifetime.
The temperature upper limit consists of two temperatures. The lower temperature is the maximum isotherm temperature on which the column can be used normally. The higher temperature is the maximum programming temperature and should not be held for more than 10 min.
Operating at a higher temperature than the temperature limit will reduce the column lifetime.
6. Storage
- After the experiment, cool all heated zones below 50 ℃, unload the column, seal each end of the column with cap ends or septums, put it in the column box, and store in a cool and dry place.
- Attention: Label the injection port side and detector port side, following the same direction next time for use.
Attached List — Temperature Limits of GC Columns
| Column / Phase | Temperature Limits (℃) | Column / Phase | Temperature Limits (℃) |
|---|---|---|---|
| 100% Dimethylpolysiloxane | DM-InertWax 0.25/0.32 mm | 20 to 250 | |
| DM-1 0.25/0.32 mm, 0.10–0.50 μm | -60 to 330/350 | DM-InertWax 0.53 mm | 20 to 240/250 |
| DM-1 MS | -60 to 330/350 | Modified Polyethylene glycol | |
| DM-1 0.25 mm, 1.00 μm | -60 to 330/340 | DM-FFAP 0.25/0.32 mm | 40 to 250 |
| DM-1 0.32 mm, 1.00 μm | -60 to 320/340 | DM-FFAP 0.53 mm, 0.50–1.00 μm | 40 to 240/250 |
| DM-1 0.32 mm, 1.50 μm | -60 to 310/330 | DM-FFAP 0.53 mm, 1.50 μm | 40 to 220/230 |
| DM-1 0.53 mm, 0.50–1.50 μm | -60 to 310/330 | 100% Trifluoropropylmethylpolysiloxane | |
| DM-1 0.32 mm, 3.00–4.00 μm | -60 to 280/300 | DM-200 0.25/0.32 mm, 0.25 μm | -20 to 320/340 |
| DM-1 0.32 mm, 5.00 μm | -60 to 260/280 | DM-200 MS | -20 to 320/340 |
| DM-1 HT | -60 to 400 | DM-200 0.25/0.32 mm, 0.50 μm | -20 to 310/330 |
| 5% Diphenyl-95% Dimethylpolysiloxane | DM-200 0.25/0.32/0.53 mm, 1.00 μm | -20 to 290/310 | |
| DM-5 0.25/0.32 mm, 0.10–0.50 μm | -60 to 330/350 | DM-200 0.32/0.53 mm, 1.50 μm | -20 to 280/300 |
| DM-5 0.32 mm, 1.00 μm | -60 to 330/350 | DM-200 0.53 mm, 0.50 μm | -20 to 300/320 |
| DM-5 MS | -60 to 330/350 | DM-200 0.53 mm, 3.00 μm | -20 to 260/280 |
| DM-5 0.25 mm, 1.00 μm | -60 to 320/340 | 90% Biscyanopropyl-10% Cyanopropylphenylpolysiloxane | |
| DM-5 0.32 mm, 1.50 μm | -60 to 310/330 | DM-2330 | 0 to 260/275 |
| DM-5 0.53 mm, 0.50–1.50 μm | -60 to 310/330 | Al2O3 | |
| DM-5 0.32 mm, 3.00 μm | -60 to 280/300 | DM-PLOT Alumina | 200 |
| DM-5 0.32 mm, 5.00 μm | -60 to 260/280 | DM-PLOT Alumina-Na2SO4 | 200 |
| DM-5 0.53 mm, 3.00–5.00 μm | -60 to 270/290 | DM-PLOT Alumina-KCl | 200 |
| DM-5 HT | -60 to 400 | DM-PLOT CFC | 200 |
| 5% Diphenyl-95% Phenylarylene dimethylpolysiloxane | 5A Molecular sieve | ||
| DM-5MS/LB | -60 to 330/350 | DM-PLOT MS 5A | 300 |
| 35% Diphenyl-65% Dimethylpolysiloxane | Divinylbenzene | ||
| DM-35 0.25/0.32 mm, 0.25 μm | 40 to 320 | DM-PLOT Q | 280/300 |
| DM-35 0.25/0.32 mm, 0.50 μm | 40 to 310 | DM-PLOT QS | 250 |
| DM-35 0.53 mm, 1.00 μm | 40 to 290 | DM-PLOT S | 250 |
| DM-35MS | 50 to 340/360 | DM-PLOT U | 190 |
| 50% Diphenyl-50% Dimethylpolysiloxane | Special Columns | ||
| DM-17 | 40 to 280/320 | DM-PONA | -60 to 300/340 |
| DM-17MS | 40 to 340/360 | DM-TCEP | 0 to 135 |
| 6% Cyanopropylphenyl-50% Dimethylpolysiloxane | DM-Volatile Amine | -60 to 290 | |
| DM-624 | -20 to 240 | DM-5 Amine 0.25/0.32 mm, 0.50–1.00 μm | -60 to 300/315 |
| DM-624MS | -20 to 300/320 | DM-5 Amine 0.32 mm, 1.50 μm | -60 to 290/305 |
| 14% Cyanopropylphenyl-86% Dimethylpolysiloxane | DM-5 Amine 0.53 mm, 1.00 μm | -60 to 290/305 | |
| DM-1701 0.25/0.32 mm, 0.25 μm | -20 to 280 | DM-5 Amine 0.53 mm, 3.00 μm | -60 to 280/295 |
| DM-1701 0.25/0.32 mm, 0.50 μm | -20 to 270/280 | DM-35 Amine | 0 to 220 |
| DM-1701 0.25/0.32 mm, 1.00 μm | -20 to 260/280 | DM-Wax Amine | 40 to 210/220 |
| DM-1701 0.53 mm, 0.50 μm | -20 to 260/270 | DM-TVOC | 300/320 |
| DM-1701 0.53 mm, 1.00 μm | -20 to 250/270 | DM-PAH | 40 to 340/360 |
| DM-1701 0.53 mm, 1.50 μm | -20 to 240/260 | DM-FAMEWAX | 20 to 250 |
| DM-1701 0.53 mm, 3.00 μm | -20 to 230/250 | DM-2560 | 20 to 250 |
| 50% Cyanopropylmethyl-50% Phenylmethylpolysiloxane | DM-1HT SimDist Metal | -60 to 450 | |
| DM-225 0.25/0.32 mm | 40 to 220/240 | DM-1 SimDist Metal | -60 to 430 |
| DM-225 0.53 mm | 40 to 200/220 | DM-500 SimDist Metal | -60 to 430 |
| Polyethylene glycol | DM-2887 | -60 to 360 | |
| DM-Wax 0.25/0.32/0.53 mm, 0.25–1.00 μm | 40 to 250/260 | DM-2887 Metal | -60 to 400 |
| DM-Wax 0.53 mm, 2.00 μm | 40 to 220/230 | DM-BDTG Metal | -60 to 380/430 |
Attention:
Do not exceed the phase’s maximum isotherm operating
temperature!