Lab Glassware Division 中文(Chinese)

Short Chain Amines in Water

Method: GC
Matrix: Standards
Application No.: 101852
Columns: DM-Volatile Amine 50 m x 0.32 mm
Cat No.: 8857
Conditions: Sample: 200 - 1,000 ppm Short chain amines in water
Injection: Split, 15:1, 1.0 μL, 220 ºC
Oven Temp.: 40 ºC (hold 10 min) to 250 ºC (hold 10 min) at 20 ºC/min
Carrier Gas: H2, 2.0 mL/min, 35 cm/sec, 40 ºC
Detector: FID, 250 ºC
Author: Dikma Technologies Inc.
Keyword: Short Chain Amines , GC, DM-Volatile Amine

Table: 1. Methanol
2. Dimethylamine
3. Trimethylamine
4. Isopropylamine
5. Dimethylethylamine
6. Diethylamine
7. Diethylmethylamine
8. Triethylamine